Vitalograph In2itive
High quality mobile spirometry with PC
and network connectivity
Technical Data
Product:
Hand Held Spirometer
Model Number:
2120
Subjects:
Automatic population of database from Spirotrac. Name, ID; Age, Height, Gender; Smoking status, Body Mass Index and more depending on model. Storage of up to 10,000 subjects
Environmental Data:
Selectable: Temperature (built in sensors), Barometric pressure, altitude, humidity
Test types:
Single breath tests, flow/volume loops, multi-breath testing, tidal breathing and combined VC/FVC type test methods supported
Selectable Parameters (depending on model):
VC; IVC; IC; VT (TV); TLC; RV; IRV; ERV; FRC; FVC; FIVC; FEV1; FEV3; FEV6; FVC; FEV1/VC; FEV1/FVC; FEV3/VC; FEV3/FVC; FEV1/FEV6; FEF75; FEF50; FEF25; FEF25-75; FEF25-75/FVC; FIV1; PIF, FIV1/FIVC, FIF25, FIF50, FIF75, FEF50/FIF50, FET, MVVind, FEV1 Ratio; FEV0.5; PEF L/min; PEF L/s; FEF 0.2-1.2; FEF 75-85%; FEF25%; FEF50%; FEF75%; FMFT; FIF25%; FIF50%; FIF75%; PIF; FEV0.75; Lung Age and more
Predicted Values Selectable (depending on model):
ERS/ECCS; NHANES, Polgar, Pereira, Berglund, Forsche, Gutierrez, Hedenström, Taiwan; Knudson, Crapo; Hsu; KNLW; Viljanen; SEPAR; Gulsvik; SBPT; Tamura; Ip; Quanjer; Wang; Dockery and more
Suggested Interpretation:
User selectable
Flow Technology:
Fleisch Pneumotachograph (No3 size)
Resolution:
10 mL volume; 0.01 L/s flow
Data Storage:
Stores up to 10,000 subjects and hundreds of test sessions
Accuracy when in Operating Range:
Volumes: Better than ± 3% (Max 10L / Min 0L); Flows: Better than ± 10% (Max 16L/s / Min 0.02L/s); Linearity: ± 5% in range 0.1 L/s to 16 L/s PowerSAFE™: Input 100 - 240V AC 50-60Hz, output 5V DC
Battery Pack:
Rechargeable Lithium Polymer 3.7v 2000mAH
Dimensions:
Device: 6 1/4" x 4" x 1 3/4" (with flowhead attached)
Net Weight:
Device: 8oz
Gross Weight and Size (Packed):
Weight: 3lbs 5oz, Packed Size: 8 1/2" x 4" x 6 1/4"
Storage Humidity:
10% to 95%
Storage Temperature:
0°C to 50°C
Recommended Operating Temperature Range:
17 to 37°C; Design Limits 10-40°C
Ambient Humidity:
0% to 99%
Connectivity:
USB 2
Max Test Duration:
20s FVC; 30s VC
Performance Standards:
ISO 26782:2009
ISO 23747:2007
ATS/ERS 2005
Safety Standards:
IEC 60601-1:2005
Medical Safety Standard:
Medical Devices Directive 2007/47/EC
Designed & Manufactured Under:
ISO 13485:2003
FDA 21CFR820
CMDR