Vitalograph - the difference is clear
in2itive

Vitalograph In2itive

High quality mobile spirometry with PC
and network connectivity

Technical Data

Product:

Hand Held Spirometer

Model Number:

2120

Subjects:

Automatic population of database from Spirotrac. Name, ID; Age, Height, Gender; Smoking status, Body Mass Index and more depending on model. Storage of up to 10,000 subjects

Environmental Data:

Selectable: Temperature (built in sensors), Barometric pressure, altitude, humidity

Test types:

Single breath tests, flow/volume loops, multi-breath testing, tidal breathing and combined VC/FVC type test methods supported

Selectable Parameters (depending on model):

VC; IVC; IC; VT (TV); TLC; RV; IRV; ERV; FRC; FVC; FIVC; FEV1; FEV3; FEV6; FVC; FEV1/VC; FEV1/FVC; FEV3/VC; FEV3/FVC; FEV1/FEV6; FEF75; FEF50; FEF25; FEF25-75; FEF25-75/FVC; FIV1; PIF, FIV1/FIVC, FIF25, FIF50, FIF75, FEF50/FIF50, FET, MVVind, FEV1 Ratio; FEV0.5; PEF L/min; PEF L/s; FEF 0.2-1.2; FEF 75-85%; FEF25%; FEF50%; FEF75%; FMFT; FIF25%; FIF50%; FIF75%; PIF; FEV0.75; Lung Age and more

Predicted Values Selectable (depending on model):

ERS/ECCS; NHANES, Polgar, Pereira, Berglund, Forsche, Gutierrez, Hedenström, Taiwan; Knudson, Crapo; Hsu; KNLW; Viljanen; SEPAR; Gulsvik; SBPT; Tamura; Ip; Quanjer; Wang; Dockery and more

Suggested Interpretation:

User selectable

Flow Technology:

Fleisch Pneumotachograph (No3 size)

Resolution:

10 mL volume; 0.01 L/s flow

Data Storage:

Stores up to 10,000 subjects and hundreds of test sessions

Accuracy when in Operating Range:

Volumes: Better than ± 3% (Max 10L / Min 0L); Flows: Better than ± 10% (Max 16L/s / Min 0.02L/s); Linearity: ± 5% in range 0.1 L/s to 16 L/s PowerSAFE™: Input 100 - 240V AC 50-60Hz, output 5V DC

Battery Pack:

Rechargeable Lithium Polymer 3.7v 2000mAH

Dimensions:

Device: 6 1/4" x 4" x 1 3/4" (with flowhead attached)

Net Weight:

Device: 8oz

Gross Weight and Size (Packed):

Weight: 3lbs 5oz, Packed Size: 8 1/2" x 4" x 6 1/4"

Storage Humidity:

10% to 95%

Storage Temperature:

0°C to 50°C

Recommended Operating Temperature Range:

17 to 37°C; Design Limits 10-40°C

Ambient Humidity:

0% to 99%

Connectivity:

USB 2

Max Test Duration:

20s FVC; 30s VC

Performance Standards:

ISO 26782:2009
ISO 23747:2007
ATS/ERS 2005

Safety Standards:

IEC 60601-1:2005

Medical Safety Standard:

Medical Devices Directive 2007/47/EC

Designed & Manufactured Under:

ISO 13485:2003
FDA 21CFR820
CMDR