Technical Data
Product:
Vitalograph 2120
Model:
2120
Size:
10 5/8" x 9 3/4" x 6 3/4" (2120 & cradle)
Weight:
41bs 1oz net
Parameters Measured:
VC, FVC, FEV.5, FEV.5/FVC%, FEV.75, FEV.75/FVC%, FEV1, FEV1/VC%, FEV1/FVC%, FEV1/PEF, FEV3, FEV3/FVC%, PEF, FEF25-75%, FEF25-75%/FVC, FEF75-85%, FEF.2-1.2, FEF25%, FEF50%, FEF75%, FMFT, FET, MVVInd, FIVC, FIVC/FVC%, PIF, FIF25%, FIF50%, FIF75%, FIF50%/ FEF50% and more, depending on variant.
Operating Temperature Range:
63 - 98 °F
Voltage:
110 - 250V~ 50/60Hz
Flow Detection Principle:
Fleisch Type Pneumotachograph
Volume Detection:
Flow Integration
Accuracy when operated in operating temperature range conditions:
Volume better than +/- 3% (up to 8L)
Flow better than +/- 5% (up to 15 L/s)
Maximum Displayed Volume:
> 8 L
Maximum Test Duration (FVC):
20 seconds
Exp Flow Impedance:
<0.1 kPa/L/s @ 14 L/s
Report:
PDF Vitalograph Reports or optional external inkjet printer
PERFORMANCE STANDARDS
Meets or exceeds ATS/ERS: 2005 Guidelines & all COPD Guidelines
Safety Standards:
EN60601-1 & 2
Designed and manufactured to following QA/GMP standards:
ISO 13485:2003
FDA 21CFR820
Manufactured by:
Vitalograph (Ireland) Ltd.
Gort Road Business Park
Ennis, Co Clare, Ireland