Peak Flow Meter & asmaPLAN+ Peak
Flow Meter
The simple, accurate way to monitor asthma
Technical Data
Product:
Peak Flow Meter asmaPLAN and Peak Flow Meter asmaPLAN+
Type of Device:
Variable orifice simple mechanical peak flow meter
Peak Flow Meter Standard:
EN13826:2003
Safety Standards:
EN ISO 60601-1:2005 -1 -2
Medical Safety Standard:
Medical Devices Directive 93/42/EEC 1993
Designed & Manufactured Under:
ISO 13485:2003, FDA 21CFR820, CMDR & JPAL
Range: (L/min):
50-800L/min
Accuracy:
Better than 10%
Repeatability:
Better than 5% or 10L/min whichever is greater
Inter-Instrument Variation:
Better than 5% or 10L/min whichever is greater
Scale Intervals:
10L/min below 700, 20L/min above
Daily Record Chart:
Included
Measuring Principle:
Piston/spring and moveable pointer
Frequency Response:
A/B difference less than 15L/min/15% (Annex B of EN13826)
User Instructions:
Multi-language instruction sheet
Clinic Use:
Certify accuracy at least annually using a certified calibrated spirometer.
Home Use:
Replace every three years
Net Weight:
2.6 oz net
Highest Resistance to Flow:
0.00384KPa/L/min@ 720L/min
Storage:
Temperature: 62 - 99° F
Relative Humidity:
10%-95%
Dimensions:
5 1/4" x 2 1/4" x 1 3/8" (incl. integral mouthpiece)