Vitalograph - the difference is clear
Bacterial Viral Filters

Bacterial Viral Filters

For Effective Infection Control

Technical Data

Product:

BVF™ Bacterial Viral Filter

Model Number:

2820

Cross Contamination Efficiency Bacteria:

99.99% @ 55L/min

Cross Contamination Efficiency Virus:

99.99% @ 55L/min

Cross Contamination Efficiency Bacteria:

99.99% @ 750L/min

Cross Contamination Efficiency Virus:

99.99% @ 750L/min

Filtration Efficiency Bacteria:

99.5% @ 55 L/min

Filtration Efficiency Virus:

99.5% @ 55 L/min

Filtration Efficiency Bacteria:

99.4% @ 750 L/min

Filtration Efficiency Virus:

99.4% @ 750 L/min

Validation using Bacillus Subtillis var niger
mean dia.:

0.7 to 1 µm

Resistance to Airflow @ 750L/min (BVF alone):

0.08kPa/L/s (0.8 cmH2O)

Resistance to Airflow @ 750L/min (BVF + Fleisch):

0.12kPa/L/s (1.2 cmH2O)

Effective Deadspace (excluding Tapers):

65mL

Material:

Electrostatic nanofibre technology filter medium